How can I clean and disinfect the reusable tips of my HEINE Otoscope?

The reusable tips shall be cleaned and disinfected manually (in an immersion bath) or by machine. Recommended agents: Cleaning agent: Neodisher® MediClean Disinfection agent EU: alcoholic (e.g. rotasept®); USA: Cidex®OPA Solution The reusable tips can be reprocessed up to 360 cycles. Disposable tips (HEINE AllSpec/HEINE UniSpec) are for single use only.