How can I clean and disinfect the reusable tips of my HEINE Otoscope?

The reusable tips shall be cleaned and disinfected manually (in an immersion bath) or by machine. Recommended agents: Cleaning agent: Neodisher® MediClean Disinfection agent EU: alcoholic (e.g. rotasept®); USA: Cidex®OPA Solution The reusable tips can be reprocessed up to 360 cycles. Disposable tips (HEINE AllSpec/HEINE UniSpec) are for single use only.

What handles can be used with the HEINE BETA400 LED F.O. Otoscope?

The HEINE BETA400 LED F.O. Otoscope can be used with all HEINE BETA rechargeable handles.