HEINE EasyClean LED Laryngoscope Handle
100% waterproof. In the case of immersion disinfection ('High-Level Disinfection') or low temperature sterilisation processes (STERRAD®, STERIS®), neither the batteries nor the LED light unit need to be removed.
- 100 % waterproof
- Autoclavable LED technology
- Simple, safe reprocessing
- LED in HEINE Quality – or LED HQ.
Configurations
More details
Complete handle (without batteries)
100% waterproof. In the case of immersion disinfection (‘High-Level Disinfection’) or low temperature sterilisation processes (STERRAD®, STERIS®), neither the batteries nor the LED light unit need to be removed.
Autoclavable LED technology: No removal of the light unit required for use in autoclaves, just batteries.
Non-slip “Wave Design” improves the reliability of the wipe disinfectant for the handle after use (‘Intermediate Disinfection’).
HEINE has set a new standard which maintains that only the best is good enough, from the selection of materials to processing, from light intensity and thermal management to a colour rendering index as high as possible.
That’s what we call LED in HEINE Quality – or LEDHQ.
Maximum brightness: 3x brighter than conventional halogen light.
Virtually unlimited hours of working life.
Typ. 10 hours of operating time (EasyClean LED Laryngoscope Handle incontinuous operation).
Compatible with all laryngoscope blades according to ISO 7376 (Green Standard).
Product information
Instructions for use | Reprocessing
Others | Declaration of conformity
Download all (zip)
Further articles | Accessories HEINE EasyClean LED Laryngoscope Handle

114mm Overall length, 98mm Blade length, 14mm Distal width
F-000.22.102
82mm Overall length, 66mm Blade length, 9mm Distal width
F-000.22.100
90mm Overall length, 72mm Blade length, 9mm Distal widthPack of 25 pieces
F-000.22.761
76mm Overall length, 51mm Blade length, 10mm Distal width
F-000.22.119FAQ - Frequently Asked Questions
What shall I do to avoid reactions with subsequent treatment stages?
After cleaning, the handle must be rinsed free of residue in order to avoid reactions with subsequent treatment stages/ damage to the materials.
Can I perform hygienic reprocessing without removing the batteries?
If you use reprocessing programs with temperatures below 60°C (Low-Temperature Procedures) or if the battery manufacturer´s instructions allow the temperatures of the used procedures, you can reprocess the handle without removing the batteries.
For which procedures do I need to remove the batteries?
Procedures using tempperatures over 60 C , e.g.: - automated cleaning and disinfection in a washer disinfector
- autoclave/steam sterilization
For which procedures can I leave the batteries in the handle?
Procedures using tempperatures under 60°C , e.g.:
- manual cleaning and disinfection
- low temperature sterilization (VHP/STERRAD)
Can I combine automated cleaning and disinfection with STERRAD/VHP(Steris)?
Yes you can, but you need to remove the batteries before automated cleaning and disinfection.