FAQ - Frequently Asked Question

HEINE DELTA 20T Dermatoscope


General


What handles can be used with the HEINE DELTA20 T Dermatoscope?

All HEINE BETA handles can be used with the HEINE DELTA20 T Dermatoscope.


How do I change the light source in my HEINE DELTA20 T Dermatoscope?

With the HEINE DELTA20 T Dermatoscope the LED cannot be changed.


Is immersion liquid required when using the HEINE DELTA20 T Dermatoscope?

The choice to use liquid when performing Dermoscopy is a personal one. With the advent of Polarized lighting technology, liquids are not required, but their use does provides additional diagnostic information.


Hygienic Reprocessing


How does the Toggle Function of the HEINE DELTA20 T Dermatoscope work?

The DELTA20 T offers 2 illumination modes – Polarized and Non-Polarized. When the instrument is initially turned on it will always be in Polarized mode. To change to Non-Polarized, simply press the black button on either side of the instrument “neck”. To return to Polarized mode, press the button again. If at any time you are unsure what mode you are in, simply turn the instrument OFF and then back ON and it will default to the Polarized mode.